Model Number IRC5PO2V |
Device Problems
Loss of Power (1475); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional information become available for the patient a supplemental record will be filed.
|
|
Event Description
|
Provider states the unit has an error code of 1 red 3 green, shutting off after a few minutes.Provider states upon start up the unit does not alarm.
|
|
Manufacturer Narrative
|
Additional/updated information was added to reflect the compressor being returned to the manufacturer for evaluation.The result of the evaluation was that the compressor was noisy.However, that is not enough information to confirm the reported problem of the unit having a 1 red, 3 green error code.The power switch will not be returned for evaluation, so the reported problem of the unit not alarming upon startup will not be confirmed.
|
|
Event Description
|
Provider states the unit has an error code of 1 red 3 green, shutting off after a few minutes.Provider states upon start up the unit does not alarm.
|
|