MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM; EXTERNAL COMPONENT Back to Search Results
Model Number DACAPO POWERPACK (TEMPO ACCU WYON)
Device Problems Break (1069); Fluid/Blood Leak (1250); Chemical Spillage (2894)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Conclusion: the returned device was visually inspected upon arrival.A mechanically damaged housing was observed.The observed damage did not allow electrical testing to be conducted.The damage to the battery housing was clearly the reason for the malfunction of the device returned by the end-user.This is a combined initial and final report.
 
Event Description
Returned dacapo power pack showed a broken housing with electrolyte leakage.However, neither patient contact to battery chemistry nor any injuries were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Type of Device
EXTERNAL COMPONENT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key5392984
MDR Text Key37233752
Report Number9710014-2016-00043
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 01/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDACAPO POWERPACK (TEMPO ACCU WYON)
Device Lot Number0911D028
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received01/19/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age9 YR
-
-