DEPUY SYNTHES POWER TOOLS BASIC CONSOLE FOR ELECTRIC PEN DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.002 |
Device Problem
Failure to Shut Off (2939)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported that the console device did not shut off and needed to be unplugged to turn the unit off.The event was not reported to have occurred during surgery.It was not reported if there were any delays to a scheduled surgical procedure, or if a spare device was available for use.There were no patient or user injuries reported.It was reported there was no medical intervention or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the motor connector malfunctioned, electric pen drive started running immediately after plugging into console and would not turn off unless unplugged.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due wear from normal use and servicing.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|