DEPUY SYNTHES POWER TOOLS SMALL BATTERY DRIVE (COLIBRI); INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.001 |
Device Problem
Device Stops Intermittently (1599)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Additional narrative: the actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device had a faulty motor and plate.It was further determined that the device had a general wear and did not have a sufficient permanent tightness.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to improper cleaning and maintenance, which is user error, misuse, and / or abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure, it was discovered that the small battery drive device stopped intermittently.It was not reported if there were any delays in the surgical procedure or if a spare device was available.There was patient involvement reported.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|