• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM; MDS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM; MDS Back to Search Results
Model Number MT22430-PNK
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/11/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Patient's mother contacted dexcom on (b)(6) 2016, to report that on (b)(6) 2016, their receiver's display screen became frozen.There was no report of injury or medical intervention.No additional event or patient information is available.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The complaint device was returned for evaluation.The device was visually inspected and no defect was found.Functional testing was performed and there was no failure detected related to the complaint.The receiver log was reviewed and no errors were observed.The reported event of a frozen display screen was not confirmed.A root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
MDS
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key5414445
MDR Text Key37744340
Report Number3004753838-2016-00864
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000101
UDI-Public(01)00386270000101(241)MT22430-PNK(10)5201469(17)NI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22430-PNK
Device Catalogue NumberSTR-KD-PNK
Device Lot Number5201469
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received08/26/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/04/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age5 YR
Patient Weight17
-
-