The customer complained of issues with a new lot of albumin gen.2 (alb2) reagent that doesn't seem to be maintaining calibration.The new alb2 reagent lot number is 119319 expiring 02/28/2017 received on 01/11/2016.The old lot number was 619340 and the customer is not having any issues with that lot.The new reagent lot was loaded onto the instrument on (b)(6) 2016 and the customer received calibration errors.Quality controls (qc) were in range.As the reagent pack was used up, the qc values dropped out of range.The customer repeated 35 patient samples tested for alb2.Data was provided for 30 patient samples.Of the data provided, 7 patient samples were erroneous and reported outside of the laboratory.Corrected reports were issued after repeat testing was performed.(b)(6).No adverse event occurred.The c 702 analyzer serial number (b)(4).A specific root cause could not be identified.It was noted that the reagent packs may have frozen during delivery on (b)(6) 2016 as the same reagent pack lot delivered on (b)(6) 2016 was not affected and is holding their calibration.The instrument was checked and calibration and quality controls were acceptable.
|