IRIS INTERNATIONAL ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM; AUTOMATED URINALYSIS SYSTEM Back to Search Results
Catalog Number 700-7177-001
Device Problem Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/18/2016
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) was at the customer site on (b)(6) 2016.The fse ran numerous racks of distilled water and noticed that intermittent strip jams were occurring at extractor.Fse performed wiper/extractor calibration and did not observe pads being removed by extractor.The repairs were verified per established service procedures.Bec internal identifier for this report is (b)(6).
 
Event Description
The customer reported that there were pads falling off the strips into the strip provider module (spm) on their ichem velocity automated urine chemistry system.Erroneous patient results were not reported out of the lab and there was no change or effect to patient treatment in connection to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM
Type of Device
AUTOMATED URINALYSIS SYSTEM
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer (Section G)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer Contact
gopal mohanty
9172 eton avenue
chatsworth, ca, CA 91311
8185277379
MDR Report Key5435687
MDR Text Key38228520
Report Number2023446-2016-00080
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number700-7177-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/18/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/13/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown