• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERGE HEALTHCARE MERGE HEMODYNAMICS COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERGE HEALTHCARE MERGE HEMODYNAMICS COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number MERGE HEMO V.9.30
Device Problems Device Operational Issue (2914); Patient Data Problem (3197)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/21/2013
Event Type  malfunction  
Manufacturer Narrative
Upon retrospective review, this issue was determined to be reportable as an mdr.
 
Event Description
Merge hemo monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure. The customer contacted merge on (b)(6) 2013 to report that the case is locked, and asked for help to unlock for data entry. The locks were removed from study and the reason was due to 3 labs being locked into the study. Although it was not showing open on any station, the lock still persisted. On (b)(4) 2013, merge corrected tables in sql to unlock study. The system is designed to allow users to lock a study or tab within the study to make changes. This ensures that two users can not make changes to the same study at the same time. However, in some instances the system will lock tabs within a study even when a second user does not have the study open on a different workstation. This "phantom" lock mimics the behavior described above except the lock is not caused by a second user but by the sql table. The lock might prevent users from calculating certain calculation that are required during the procedure since height and weight might be unavailable. (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMERGE HEMODYNAMICS
Type of DeviceCOMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge dr.
hartland, WI 53029
2629123570
MDR Report Key5437723
MDR Text Key38228006
Report Number2183926-2015-00368
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation
Remedial Action Inspection
Type of Report Initial
Report Date 02/15/2016
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/16/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberMERGE HEMO V.9.30
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/21/2013
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/10/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage
Removal/Correction Number2183926-02/15/2016-048-C

-
-