Efilm workstation is a software application that is used for viewing medical images.Efilm workstation receives digital images and data from various sources (including but not limited to ct, mr, us, rf units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources).A customer reported ((b)(4)) that after setting up v 4.0.3 of the efilm server, they were unable to receive images unless efilm service is restarted.This was determined to be a potential safety issue in the event the information is not available for review.The investigation discovered that images were being stored on a network share.Merge informed the customer that the system does not support network storage.
|