Model Number MT22495
Device Problems Self-Activation or Keying (1557); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/20/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Patient contacted dexcom on (b)(6) 2016 to report the receiver initialized without a manual restart on (b)(6) 2016.There was no report injury or medical intervention.No additional patient or event information is available.Data download log was provided by the patient and reviewed for evaluation on (b)(6) 2016.Complaint for initializing screen without a manual restart could not be confirmed.The root cause could not be determined post investigation.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The complaint device was returned for evaluation.The device was visually inspected and no defect was found.Functional testing was performed and there was no failure detected related to the complaint.The receiver log was reviewed and no errors were observed.The reported event of initializing screen without manual restart was not confirmed.A root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM
Type of Device
MDS
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key5439171
MDR Text Key38746491
Report Number3004753838-2016-05752
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000156
UDI-Public(01)00386270000156(241)MT22495(10)5204960(17)NI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22495
Device Catalogue NumberSTR-DR-001
Device Lot Number5204960
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received03/15/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age65 YR
-
-