A medtronic representative, following-up, reported the target was changed but the entry was re-set as well.Pointmerge was used for registration, the 3d model looked perfect.The fiducial placement was mildly symmetrical.The entry point was a few millimeters off the surface.On 02/12/2016 software analysis was unable to determine probable cause with the information provided.It is suspected that moving the entry point may have caused the reported behavior.No parts have been received by manufacturer for analysis.No further issues have been reported.
|
A medtronic representative reported that, while in a cranial biopsy procedure, it was alleged that the first pass with the needle went beyond target less than 5 millimeters.Entry was re-set and passed again without issue.The surgeon completed the procedure with the use of the navigation system.There was no delay of therapy.There was no impact on patient outcome.
|