Model Number PM3242
Device Problem Difficult to Remove (1528)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/22/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during a system upgrade procedure, the right ventricular lead could not be removed from the header port of the pulse generator.Additional force was applied and the lead was successfully removed.The device was explanted and replaced as planned.
 
Manufacturer Narrative
Final analysis confirmed that it was difficult to insert the test lead(s) into the pulse generator's header and the lead insertion force used to insert the lead was out of specification for the product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE QUADRA RF CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
15900 valley view court
sylmar, CA 91342
8184934022
MDR Report Key5442588
MDR Text Key38431053
Report Number2017865-2016-00965
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/17/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2015
Device Model NumberPM3242
Device Lot Number4573799
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/17/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR
-
-