CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problem Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/26/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the physician's tablet had been charging for over 24 hours with the orange/red light visible on the tablet, but the tablet would not power on when the power button was pressed.Troubleshooting did not resolve the issue.A new tablet was provided to the physician and the faulty tablet was received for analysis.Analysis is underway, but has not been completed to date.
 
Event Description
Analysis of the tablet was completed on 03/08/2016.The cause for the anomaly is associated with an upside down power button.Once the power button was installed correctly, no further anomalies associated with the tablet were identified during the analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5450507
MDR Text Key39206757
Report Number1644487-2016-00373
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received03/08/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/27/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-