STRYKER ORTHOPAEDICS-MAHWAH HEADED PIN IMPACTOR EXTRACTOR; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 6541-4-300
Device Problems Degraded (1153); Flaked (1246); Material Integrity Problem (2978)
Patient Problem No Patient Involvement (2645)
Event Date 01/29/2016
Event Type  malfunction  
Manufacturer Narrative
When completed, the investigation results will be submitted in a supplemental report.
 
Event Description
On (b)(6), myself and one of the sterile processing people noticed that the instruments were starting to chip and flake off.The rubber handles were flaking off, due to so many uses.
 
Manufacturer Narrative
An event regarding santoprene handle degradation was reported.The event was confirmed.Method and results: device evaluation and results: visual analysis confirmed the reported event.The handle was returned degraded.Medical records received and evaluation: not performed because patient factors did not contribute to the reported event.Device history review: review of the device history records indicates that all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there has been two other events for the lot referenced.Conclusions: capa was initiated because a series of complaints were received for triathlon instrumentation where green santoprene over-mold handles have been reported to be degrading, becoming sticky and unstable.This event was determined to be under the scope of the capa.Chemical and cytotoxicity tests showed the degraded santoprene is non-toxic.The ability to clean and sterilize the degraded instruments has not been compromised.
 
Event Description
On (b)(6) myself and one of the sterile processing people noticed that the instruments were starting to chip and flake off.The rubber handles were flaking off, due to so many uses.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEADED PIN IMPACTOR EXTRACTOR
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5457783
MDR Text Key38979306
Report Number0002249697-2016-00558
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123486
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number6541-4-300
Device Lot NumberN3L58
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received03/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/23/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-