• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. DISCOVERY ELBOW; DISC CONDYLE KIT WITH HEXALOBULA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. DISCOVERY ELBOW; DISC CONDYLE KIT WITH HEXALOBULA Back to Search Results
Catalog Number 114700
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Code Available (3191)
Event Date 01/25/2016
Event Type  Injury  
Event Description
Revision surgery - due to the patient's implants being loose.
 
Manufacturer Narrative
The reason for this revision surgery was loosening.The length of the in-vivo service is unknown since the original surgery was not provided or determined.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing history of this part.All parts were found to meet design and manufacturing specifications.No non-conforming material reports (ncmr's) were associated with this part.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed to be non-product related.The surgeon reported no issues associated with the explanted product and provided no information that definitively described the root cause or reason of the elbow looseness.The scope of this investigation is limited without having the parts available to djo surgical for evaluation.Factors that may contribute to joint instability that are not associated with the implants are: improper implant selection, degenerative bone disease, patient non-compliance with medical instructions.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISCOVERY ELBOW
Type of Device
DISC CONDYLE KIT WITH HEXALOBULA
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
teffany hutto
9800 metric blvd.
austin, TX 78758-5445
5128346255
MDR Report Key5458523
MDR Text Key38963947
Report Number1644408-2016-00068
Device Sequence Number1
Product Code JDC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051975
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Catalogue Number114700
Device Lot Number120920
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/08/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-