Model Number MERGE CARDIO 8.30
Device Problem Data Problem (3196)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/05/2013
Event Type  malfunction  
Manufacturer Narrative
Upon retrospective review, this issue was determined to be reportable as an mdr.
 
Event Description
Merge cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information including images, hemodynamic studies and reports, measurements (via import from dicom structured reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data.Customer reported that measurements are not populating in the report.There were 4 images that have measurements on them, however the measurements are not populating from the images from the report.Technical support determined that the sequia modalities should not send in sr files.This resulted in a lot of unmapped measurements that were proprietory measurements sent from the modality.Technical support asked for a test patient in order to map some the measurements to meet the modality.The site later reported that the modality is no longer being used.There were no reports of patient injury.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE CARDIO
Type of Device
PICTURE ARCHIVING AND COMMUNICATIONS SYSTEM
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge dr.
hartland, WI 53029
2629123570
MDR Report Key5476069
MDR Text Key39531615
Report Number2183926-2016-00110
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/05/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/03/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE CARDIO 8.30
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/05/2013
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-