Model Number 1020000
Device Problem Smoking (1585)
Patient Problem Burn(s) (1757)
Event Date 12/10/2015
Event Type  Injury  
Event Description
The manufacturer received information alleging a patient was smoking a cigarette while using an everflo oxygen concentrator.The patient received second degree burns and was hospitalized.The fire marshall confirmed the cause of the fire was the patient smoking a cigarette while using the device.The device was evaluated by a third party service center.The device had no evidence of thermal damage to any internal components.The exterior of the device had evidence of thermal damage caused by an external source.The device was repaired and placed back into service.Product labeling states, "do not smoke, allow others to smoke or have open flames near the concentrator when it is in use.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLO OXYGEN CONCENTRATOR
Type of Device
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS INC
175 chastain meadows court
kennesaw GA 30144
Manufacturer (Section G)
RESPIRONICS INC
175 chastain meadows court
kennesaw GA 30144
Manufacturer Contact
wilfredo alvarez
312 alvin drive
new kensington, PA 15068
7243349303
MDR Report Key5477668
MDR Text Key39607544
Report Number1040777-2016-00010
Device Sequence Number1
Product Code CAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061261
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 02/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/03/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number1020000
Device Catalogue Number1020000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received02/11/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/26/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age66 YR
-
-