Model Number 6382 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/11/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).This flow sensor was on a delphin centrifugal system.The field service representative (fsr) verified the intermittent operation of the flow sensor.The fsr installed and tested a new flow sensor.The unit operated to manufacturer specifications and was returned to clinical use.The suspect part was returned to the manufacturer for further evaluation.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the flow sensor had failed.The user facility's biomedical engineer (biomed) reported the flow sensor initially worked during setup testing, but then the flow readings stopped.This was resolved by installing a flow sensor from another system.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
Per follow-up with the field service representative (fsr) on (b)(6) 2016: the circuit was primed and tubing in the flow sensor.Flow displayed and then went to 0.0.The flow sensor was located between oxygenator and arterial filter on the circuit.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed by the field service representative (fsr).During the laboratory evaluation, there was no loss of flow readings occurred during lab testing.The product surveillance technician (pst) operated the flow sensor over a two day period with no dropout in flow readings.The pst agitated the flow sensor¿s head, connector and cable with no loss of flow readings.The pst measured signal strength from the flow sensor and it was 76%.Nothing was observed that would cause failure.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a follow-up emdr will be filed accordingly.
|
|
Search Alerts/Recalls
|