TERUMO CARDIOVASCULAR SYSTEMS CORP. 6 INCH ROLLER PUMP FOR ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP, SYSTEM 1 Back to Search Results
Model Number 801041
Device Problems Fluid/Blood Leak (1250); Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 02/11/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This complaint is related to emdr #1828100-2016-00149.During the laboratory evaluation, the product surveillance technician (pst) initiated rotation of the roller pump with no tubing installed.No problem was found as the device operated through entire speed range.The pst performed jam testing and the roller pump passes by correctly displaying "pump jam." he repeated jam testing multiple times and occasionally got ¿underspeed¿ prior to ¿pump jam,¿ this is typical.No problem found.The pst began rotation at optimal occlusion (speed not recorded).After approx.1 hour, a ¿beltslip¿ error message was seen briefly on the local display.It's possible the likely cause of this issue.The product will be sent to service to be brought to manufacturers specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a follow-up emdr will be filed accordingly.
 
Event Description
The service repair technician (srt) reported that during routine testing of the device at the service center, a "pump jam" error intermittently activates with no load on the roller pump.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.Service records show the bearing has not been replaced since initial manufacture, indicating the pump still had the generation 1 bearing.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6 INCH ROLLER PUMP FOR ADVANCED PERFUSION SYSTEM 1
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP, SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5480989
MDR Text Key39931219
Report Number1828100-2016-00150
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131618
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number801041
Device Catalogue Number801041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received04/04/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/09/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number1828100-05/04/12-010-C
Patient Sequence Number1
-
-