TERUMO CARDIOVASCULAR SYSTEMS CORP. 6 INCH ROLLER PUMP FOR ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP, SYSTEM 1
|
Back to Search Results |
|
Model Number 801041 |
Device Problems
Fluid/Blood Leak (1250); Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/11/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).This complaint is related to emdr #1828100-2016-00149.During the laboratory evaluation, the product surveillance technician (pst) initiated rotation of the roller pump with no tubing installed.No problem was found as the device operated through entire speed range.The pst performed jam testing and the roller pump passes by correctly displaying "pump jam." he repeated jam testing multiple times and occasionally got ¿underspeed¿ prior to ¿pump jam,¿ this is typical.No problem found.The pst began rotation at optimal occlusion (speed not recorded).After approx.1 hour, a ¿beltslip¿ error message was seen briefly on the local display.It's possible the likely cause of this issue.The product will be sent to service to be brought to manufacturers specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a follow-up emdr will be filed accordingly.
|
|
Event Description
|
The service repair technician (srt) reported that during routine testing of the device at the service center, a "pump jam" error intermittently activates with no load on the roller pump.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.Service records show the bearing has not been replaced since initial manufacture, indicating the pump still had the generation 1 bearing.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|