• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS DUAL COOLER HEATER CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS DUAL COOLER HEATER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS DUAL COOLER HEATER CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS DUAL COOLER HEATER Back to Search Results
Model Number 11160
Device Problem Insufficient Information (3190)
Patient Problem No Information (3190)
Event Type  No Answer Provided  
Manufacturer Narrative
(b)(4). This complaint is related to emdr #1828100-2016-00151, #1828100-2016-00152 and #1828100-2016-00153. Per the product surveillance supervisor, there will be no surgery related or customer information as these complaints are not for a specific event. Diligence for additional information cannot be performed. This complaint is not verifiable. There has been no complaints reported to the manufacturer linking the dual cooler heater (dch) unit to nontuberculous mycobaterium (ntm) infections. The manufacturer provides established requirements for sanitization and instructions for correctly maintaining the system. These are clearly outlined in the dch operators manual revision b and dch cleaning guide revision c. Included in the requirements is a daily sanitation procedure with a check of the chlorine levels, weekly cleaning and sanitation and a semiannual decaling procedure. Decontamination is required whenever biofilm is evident in the system (discoloration or cloudiness in the water system). The dch is beyond service life at this time. If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly. (b)(4).
 
Event Description
A u. S. Food and drug administration (fda) safety communication was issued for nontuberculous mycobaterium (ntm) infections associated with heater cooler devices. This complaint is specific to dual cooler heater device. This complaint will also include a journal article from centers for disease control and prevention on 'transmission of mycobacterium chimaera from heater-cooler units during cardiac surgery despite an ultraclean air ventiliation system' that was released ahead of print for the june 2016 issue (volume 22, number 6). There was no patient involvement.
 
Manufacturer Narrative
It was found that the sanitizer listed in the cleaning guide and instructions for use (ifu) was obsolete. Replacement sanitizer has been identified and the ifu and cleaning guide will be updated. If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Manufacturer Narrative
Per fda, the potential root causes of contamination from heater-cooler devices includes ntm bacteria transmitted through the air (aerosolization), laminar flow disruption and the heater-cooler design. If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSARNS DUAL COOLER HEATER
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS DUAL COOLER HEATER
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5484342
MDR Text Key39981294
Report Number1828100-2016-00154
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PREAMENDMEN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 05/18/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number11160
Device Catalogue Number11160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/12/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-