• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number 250
Device Problems Break (1069); Failure to Interrogate (1332)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/08/2016
Event Type  malfunction  
Event Description
It was reported the physician was having a difficult time using his programming wand.It was later reported the company representative was having trouble interrogating a patient with the physician's handheld device.Troubleshooting was performed and it was found the battery cover on the wand was broken and originally thought to be the issue; however, this was ruled out as the physician had used the same handheld device with a new wand earlier in the date.Additional troubleshooting was performed and showed handheld device was able to be charged using a known working power cord.The handheld device was working properly after being charged.The faulty power cable was received by the manufacture.Analysis is expected; however, it has not been completed to date.
 
Event Description
Product analysis for the returned ac power adapter cabled was completed.Analysis was performed on the ac adapter; however, no anomalies associated with the ac adapter performance were noted during testing.The ac adapter performed according to functional specifications.The programming wand was reported to be lost and is not expected to be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5486193
MDR Text Key40139257
Report Number1644487-2016-00496
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number250
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received03/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age52 YR
-
-