• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number T167
Device Problems Over-Sensing (1438); Pacing Problem (1439); Ambient Noise Problem (2877)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/27/2009
Event Type  Injury  
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.An initial report was previously submitted to fda on 05/07/2009; however due to emdr submission issue related to the follow-up report, this is being filed again as an initial report.
 
Event Description
This device was explanted for reported normal battery depletion approximately seven years later.The device was returned for analysis.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALITY 2
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5505013
MDR Text Key40500421
Report Number2124215-2016-04363
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 01/27/2009
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/17/2009
Device Model NumberT167
Other Device ID NumberVITALITY 2 DR EL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received01/27/2009
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/19/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
1882TC
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age67 YR
-
-