• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number H7493926028400
Device Problem Occlusion Within Device (1423)
Patient Problems Chest Pain (1776); Myocardial Infarction (1969); ST Segment Elevation (2059); Thrombosis (2100); Ventricular Fibrillation (2130); Vomiting (2144); Cardiogenic Shock (2262); Respiratory Failure (2484); Complete Heart Block (2627)
Event Date 02/17/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).Device is a combination product.Device evaluated by mfr: it is indicated that the device will not be returned for evaluation.A review of the batch history, historical trending, and similar complaint trending review for the product family will be conducted.If there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Event Description
Same case as mdr id: 2134265-2016-01694, 2134265-2016-02196, and 2134265-2016-02197.It was reported that stent thrombosis occurred.The target lesion was located in the right coronary artery (rca).An optical coherence tomography (oct) guiding catheter was used to deploy four synergy ii drug-eluting stents sized 4.00 x 20, 4.00 x 28, 4.00 x 24, and 4.00 x 8.The procedure was completed with very good results.The patient was sent to the intensive care unit (icu) and within an hour and a half, the patient went into ventricular fibrillation, developed cardiogenic shock and respiratory failure, and was subsequently sent back to the catheter laboratory.It was noted that the implanted synergy stents had clotted off.A non-bsc extraction catheter and angiojet thrombectomy catheter were then advanced to extract the clot, and a temporary pacer and a non-bsc ventricular assistive device was placed for support.The procedure was completed with no further patient complications reported, but the patient's status was critical post procedure.The patient remains hospitalized and in stable status with chronic kidney injury.
 
Manufacturer Narrative
(b)(4).
 
Event Description
It was further reported that the patient presented to the emergency room with a 6 day history of left sided upper chest pain with worsening in frequency and intensity, diagnosed as unstable angina.The index procedure was performed on the same day the patient presented to the emergency room.During the index procedure, prior to stent placement, two passes of aspirational thrombectomy with a non-bsc catheter were performed, yielding a small amount of dark red thrombus.The mid and distal right coronary artery was treated with multiple inflations from 3 different non-bsc balloon catheters.After initial angioplasty, the mid and distal rca are diffusely diseased with long areas of 70-80% stenosis in the mid vessel and focal 70-80% in the distal vessel.The stent balloons were used to perform post dilatation of the stents and the overlap zones in multiple inflations p to 16 atms.The physician was unable to pass longer stents due to the tortuosity, so multiple short stents had to be used.There was evidence of st elevation with no reflow distally after opening the artery.This was treated with intracoronary nitroglycerin.At the time of the event, the patient vomited and received acls, defibrillation and intubation at bedside.The patient was in complete heart block and receiving dopamine an dobutamine upon arrival to the cath lab.The aspiration thrombectomy catheters removed a large amount of red and while thrombus, even more than removed earlier in the day.A small amount of residual thrombus was left in the distal rca.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Type of Device
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5505052
MDR Text Key40522805
Report Number2134265-2016-01695
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/11/2016
Device Model NumberH7493926028400
Device Catalogue Number39260-2840
Device Lot Number18345902
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/18/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/11/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age58 YR
-
-