• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION AED PRO; DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION AED PRO; DEFIBRILLATOR Back to Search Results
Model Number AED PRO
Device Problems Device Displays Incorrect Message (2591); Communication or Transmission Problem (2896)
Patient Problem Death (1802)
Event Date 02/16/2015
Event Type  Death  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that while attempting to treat a patient (age & gender unknown) the device was unable to obtain an ecg signal via electrode pads and prompted "check pads".Complainant indicated that the patient expired.
 
Manufacturer Narrative
This follow up medwatch report is reporting the evaluation of the device.This follow up medwatch report is also correcting information submitted on the initial medwatch report.The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a faulty transistor on the flex circuit assembly.Analysis for reports of this type has not identified an increase in trend.
 
Event Description
Complainant alleged that while attempting to treat a patient (age & gender unknown) the device was unable to obtain an ecg signal via electrode pads and the device inappropriately displayed an "attach pads" message.Complainant indicated that the patient expired.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AED PRO
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key5522580
MDR Text Key41040366
Report Number1220908-2016-00824
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2016
Is this an Adverse Event Report? Yes
Device Operator Other
Device Model NumberAED PRO
Device Catalogue NumberAED PRO
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received03/18/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death;
-
-