Model Number AED PRO |
Device Problems
Device Displays Incorrect Message (2591); Communication or Transmission Problem (2896)
|
Patient Problem
Death (1802)
|
Event Date 02/16/2015 |
Event Type
Death
|
Manufacturer Narrative
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that while attempting to treat a patient (age & gender unknown) the device was unable to obtain an ecg signal via electrode pads and prompted "check pads".Complainant indicated that the patient expired.
|
|
Manufacturer Narrative
|
This follow up medwatch report is reporting the evaluation of the device.This follow up medwatch report is also correcting information submitted on the initial medwatch report.The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a faulty transistor on the flex circuit assembly.Analysis for reports of this type has not identified an increase in trend.
|
|
Event Description
|
Complainant alleged that while attempting to treat a patient (age & gender unknown) the device was unable to obtain an ecg signal via electrode pads and the device inappropriately displayed an "attach pads" message.Complainant indicated that the patient expired.
|
|