It was reported that while hospitalized due to a non-device related issue, the patient developed episodes of ventricular tachycardia, ventricular fibrillation.These episodes were observed on stored electrograms after pulse generator interrogation.For some of the episodes the patient was cardioverted externally.It was noted that all episodes started at the same hourly timestamp.On (b)(6) 2016 cap confirm was turned off and no new episodes of vt/vf were seen after reprogramming.There were no further consequences for the patient.
|