• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE INSULIN INFUSION PUMP Back to Search Results
Device Problem Failure to Prime (1492)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The pump has not been returned to animas for evaluation. If the device is returned, an evaluation shall be completed and a supplemental report will be filed. No conclusions can be made at this time.
 
Event Description
On (b)(6) 2016, the reporter contacted animas, alleging a prime (loss of prime) issue. It was reported that the loss of prime occurred 3 or more times. This complaint is being reported because the reported issue was not resolved with troubleshooting. There was no indication that the product caused or contributed to an adverse event.
 
Manufacturer Narrative
Follow-up # 1 date of submission 5/03/2016 device evaluation: the device has been returned and evaluated by product analysis on 4/18/2016 with the following findings: there was no evidence of loss of prime errors, alarms or warnings observed in the black box history. The pump successfully completed the rewind, load cartridge and prime steps with no alarms. The force sensor calibration passed within specification. The pump was exercised for 24 hours with no loss of primes occurring therefore the issue was not duplicated. The battery cap and cartridge cap were not returned with the pump and test caps were used to complete the investigation. (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameANIMAS VIBE
Type of DeviceINSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key5544346
MDR Text Key42080934
Report Number2531779-2016-07087
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 03/22/2016
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/04/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2016
Was the Report Sent to FDA? Yes
Device Age7 MO
Event Location No Information
Date Manufacturer Received03/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/21/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-