"unstable patient admitted with very high white blood cell count (>200).
A long procedure ensued, including pericarditis, valve replacement and ao replacement (lots of vasopressors given) which ended up with a 5 hour bypass time.
At 3 hours the oxygenator began to fail, so the fio2 was increased to 100% but the po2 was still dropping.
They decided to swap out the oxygenator at this point.
Suspected clogging of the fibers due to leucocytosis, with rapid lipid dropout from the patient.
" (b)(4).
|
The oxygenator was returned to maquet cardiopulmonary (b)(5) for evaluation.
The oxygenator was investigated in the decontamination/complaints laboratory.
On the blood inlet side and on the blood outlet side, no clots were visible, and none were flushed out.
The oxygenator was cleaned with sodium hypochlorite.
No leakage was found.
The oxygenator was sent to the qa laboratory for performance tests.
In the qa-laboratory, the oxygenator was tested for its gas (o2 and co2) exchange and pressure drop performance (delta p measurement) according to lv009 with bovine blood from 2017-01-23 (test date was also 2017-01-23).
The results of the tests showed that the gas exchange performance is within specified limits, but the product failed the pressure drop test.
The oxygenator was then returned to the decontamination/complaints laboratory.
All four sides of the oxygenator were removed.
The filter mats were not damaged or otherwise marked or discolored.
No other anomalies were found.
There were 74 filter mats on the gas side, and this is within specification (a count of 72, 74, or 76 is specified).
The therapy application manager was consulted for a clinical assessment of the case.
The clinical staff from (b)(6) suspected clogging of the fibers due to leukocytosis.
Despite the fact that there are no details available about blood parameters, the results from the performance tests from the qa laboratory supports the customer's hypothesis that the higher than expected pressure drop could be a result of leukocytosis, and this is the most likely reason for the reduced pressure drop performance.
A dhr review found no issues.
|