• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM; MDS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM; MDS Back to Search Results
Model Number 9500-27
Device Problems Device Operates Differently Than Expected (2913); Device Handling Problem (3265)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/09/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Patient's mother contacted dexcom on (b)(6) 2016 to report continuous glucose monitoring (cgm) inaccuracies compared to blood glucose (bg) meter that occurred on (b)(6) 2016.The sensor was inserted on (b)(6) 2016.It was reported the sensor was inserted in the arm.At the time of contact, no injury or medical intervention was reported.The complaint device was not returned for evaluation and data was not provided; therefore, the reported complaint of inaccuracies cannot be confirmed.It was reported the sensor was inserted in the arm.The dexcom g5 mobile continuous glucose monitoring system states: do not insert the sensor component of the dexcom g5 mobile system in a site other than the belly/abdomen (ages 2 years and older) or the upper buttocks (ages 2 to 17 years).The placement and insertion of the sensor component of the dexcom g5 mobile system is not approved for other sites.If placed in other areas, the dexcom g5 mobile system may not function properly.However, a root cause for the reported inaccuracy cannot be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
MDS
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key5545474
MDR Text Key41821053
Report Number3004753838-2016-11469
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000019
UDI-Public(01)00386270000019(241)9500-27(10)NI(17)NI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/04/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number9500-27
Device Catalogue NumberSTS-GL-011
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/09/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age4 YR
-
-