The lead was returned for analysis in six pieces.The overall appearance of the lead is consistent with patient manipulation of the implanted device, a twiddler.The outer silicone tubing has a compress/twisted appearance along the lead bod.Three sets of setscrew marks were seen on the connector pin providing evidence that proper contact between the setscrew and the connector pin existed at least once.A break (fracture) was identified in the positive coil and appears to have pitting at the broken end.Abraded openings were identified on both the outer and inner silicone tubing in several locations including at the break location.The connector pin also showed evidence of pitting.The closest electrode to the bifurcation has creases on the electrode ribbon most likely caused during manipulation of the electrode at implant/explant.The furthest electrode to the bifurcation has creases on the electrode ribbon and partial detachment of the electrode ribbon from the silicone helix, most likely caused during manipulation of the electrode at implant/explant.The lead assembly has dried remnants of what appear to have once been body fluids inside the inner and the outer silicone tubing.There were no performance or any other type of adverse conditions found with the pulse generator.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.The battery voltage was measured to be 3.016 volts and shows an ifi=no condition.Review of the data downloaded from the pulse generator shows an indication of increased impedance from a value of 6,392 ohms to 10,777 ohms on (b)(6) 2016 indicating that the high impedance condition was present at least as early as this date.Various electrical loads were attached to the pulse generator and results of diagnostic tests demonstrate that accurate resistance measurements were obtained in all instances.
|