A maquet field service technician (fst) evaluated the device and found some cracks on the plastic sterilizable handle support, indicating that the three holding pins got torn out while the light was manipulated.The damaged handle assembly was replaced with a new one and the device returned to service.Maquet determined that the event was caused by an excessive mechanical effort that stressed the handle support during use.The volista user manual instructs users to check the lighthead for any damage, on a daily basis.
|