• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS VOLISTA; LIGHT, SURGICAL, CEILING MOUNTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS VOLISTA; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number VLT400DFTKAIMSTP
Device Problem Device Slipped (1584)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/10/2016
Event Type  malfunction  
Manufacturer Narrative
A maquet field service technician (fst) evaluated the device and found some cracks on the plastic sterilizable handle support, indicating that the three holding pins got torn out while the light was manipulated.The damaged handle assembly was replaced with a new one and the device returned to service.Maquet determined that the event was caused by an excessive mechanical effort that stressed the handle support during use.The volista user manual instructs users to check the lighthead for any damage, on a daily basis.
 
Event Description
The customer reported that the handle assembly is not holding on and comes off easily from the central support, with components falling off.No injuries were reported.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOLISTA
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
orléans cedex 2
FR 
Manufacturer (Section G)
MAQUET SAS
orléans cedex 2
FR  
orléans cedex 2 
MDR Report Key5551488
MDR Text Key42533365
Report Number9710055-2016-00024
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K130513
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVLT400DFTKAIMSTP
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/10/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-