Model Number 5392
Device Problems No Device Output (1435); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/04/2016
Event Type  Injury  
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: analysis confirmed the customer comment of unable to pace appropriately and it was attributed to the main printed circuit board being out of specification in an electrical manner.It was also noted that the lower case battery red and black wires were pinched, though the insulation was not compromised and that one case screw was contaminated.All found defective parts were replaced and all other identified issues were resolved.(b)(4).
 
Event Description
It was reported that the external pulse generator (epg) was connected to a patient and would not pace in the demand mode.The battery was replaced and the unit was adjusted to match previous settings.When the patient was connected, it was noticed that the epg reverted to maximum output settings.The epg was returned for service.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Further analysis was performed on the main printed circuit board assembly.Visual inspection found that the main board had contamination.Bench analysis found contamination under a resistor.After the resistor was re-soldered on the pads and measured again the resistance measured correctly.Conclusion: the reported event was verified, the circuit board was contaminated under a resistor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5557832
MDR Text Key42200424
Report Number3004593495-2016-00162
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/21/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/31/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-