IPG MFG SWITZERLAND REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LNQ11
Device Problems Premature Discharge of Battery (1057); False Positive Result (1227); Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/25/2015
Event Type  malfunction  
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Recommended replacement time (rrt) was triggered on (b)(6) 2015.Daily battery trend data shows gradual rise/varying battery impedance.Early rrt alert, without corresponding battery depletion.(b)(4).
 
Event Description
It was reported that the patients implantable cardiac monitor (icm) had reached recommended replacement time (rrt) before its expected longevity.The icm remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVEAL LINQ
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
IPG MFG SWITZERLAND
route du molliau 31
tolochenaz 1131
CH  1131
Manufacturer (Section G)
IPG MFG SWITZERLAND
route du molliau 31
tolochenaz 1131
CH   1131
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5558042
MDR Text Key42794588
Report Number9614453-2016-02559
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/28/2014
Device Model NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/04/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/06/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age00059 YR
-
-