Merge unity pacs is a medical image and information management system that allows viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems.On (b)(6) 2016, merge was notified that on an unspecified date, images were skipped while scrolling with the mouse wheel.The customer reported that initially, a fracture was missed, however was caught after further investigation.Images being skipped over has the potential to lead to an incorrect and incomplete diagnosis and/or treatment, especially if the physician or health care professional is not aware that the image are being skipped.A field service engineer conducted an investigation on site and noted that there was a broken wire in the cord, causing the mouse to connect and disconnect from windows.The mouse was replaced.(b)(4).
|