MERGE HEALTHCARE MERGE UNITY PACS; PICTURE ARCHIVING AND COMMUNICATIONS SYSTEM Back to Search Results
Model Number UNITY PACS V11.0.1
Device Problems Device Issue (2379); Data Problem (3196)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/08/2016
Event Type  malfunction  
Event Description
Merge unity pacs is a medical image and information management system that allows viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems.On (b)(6) 2016, merge was notified that on an unspecified date, images were skipped while scrolling with the mouse wheel.The customer reported that initially, a fracture was missed, however was caught after further investigation.Images being skipped over has the potential to lead to an incorrect and incomplete diagnosis and/or treatment, especially if the physician or health care professional is not aware that the image are being skipped.A field service engineer conducted an investigation on site and noted that there was a broken wire in the cord, causing the mouse to connect and disconnect from windows.The mouse was replaced.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE UNITY PACS
Type of Device
PICTURE ARCHIVING AND COMMUNICATIONS SYSTEM
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge drive
hartland, WI 53029
2629123570
MDR Report Key5559977
MDR Text Key42154563
Report Number2183926-2016-00515
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K041935
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberUNITY PACS V11.0.1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/08/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-