• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP ITALIA S.R.L. - CRM FACILITY ORCHESTRA; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP ITALIA S.R.L. - CRM FACILITY ORCHESTRA; PROGRAMMER, PACEMAKER Back to Search Results
Model Number ORCHESTRA PLUS
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/23/2016
Event Type  malfunction  
Event Description
It was reported that the subject programmer was not functional.At the reboot, it remains blocked on a black screen, the led was on (green).An investigation is required.
 
Event Description
It was reported that the subject programmer was not functional.At the reboot, it remains blocked on a black screen, the led was on (green).An investigation is required.
 
Manufacturer Narrative
In-depth investigations were performed and confirmed the link between the reported issue and the observed polluted connectors found on the central processor unit board.
 
Event Description
It was reported that the subject programmer was not functional.At the reboot, it remains blocked on a black screen, the led was on (green).An investigation is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORCHESTRA
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
SORIN GROUP ITALIA S.R.L. - CRM FACILITY
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR  92140
Manufacturer (Section G)
KONTRON SUDETENSTR. 8 87600 KAUFBEUREN GERMANY
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR   92140
Manufacturer Contact
elodie vincent
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR   92140
0146013665
MDR Report Key5567646
MDR Text Key42971074
Report Number1000165971-2016-00209
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
PP980049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberORCHESTRA PLUS
Device Catalogue NumberORCHESTRA PLUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2016
Is the Reporter a Health Professional? No
Distributor Facility Aware Date03/23/2016
Date Manufacturer Received06/22/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-