Model Number ORCHESTRA PLUS |
Device Problems
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/23/2016 |
Event Type
malfunction
|
Event Description
|
It was reported that the subject programmer was not functional.At the reboot, it remains blocked on a black screen, the led was on (green).An investigation is required.
|
|
Event Description
|
It was reported that the subject programmer was not functional.At the reboot, it remains blocked on a black screen, the led was on (green).An investigation is required.
|
|
Manufacturer Narrative
|
In-depth investigations were performed and confirmed the link between the reported issue and the observed polluted connectors found on the central processor unit board.
|
|
Event Description
|
It was reported that the subject programmer was not functional.At the reboot, it remains blocked on a black screen, the led was on (green).An investigation is required.
|
|