• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED Back to Search Results
Catalog Number HAR36RR
Device Problem Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/26/2016
Event Type  malfunction  
Manufacturer Narrative
The complaint device was returned to stryker sustainability solutions for evaluation.The device was disassembled and a fracture was found in the blade.The results of the visual inspection determined that the reported event was confirmed.A review of the dhr supports that the device met all inspections and test criteria prior to release from stryker.Therefore, the most likely root causes are jaws/blade subassembly damage and incidental and prolonged activation against solid surfaces, such as bone, metal or plastic.The instructions for use state: avoid contact with any and all metal or plastic instruments or objects when the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.The reported event will continue to be monitored through post-market surveillance.
 
Event Description
It was reported that the device worked fine at the beginning of the case then suddenly stopped.The generator did not display any error messages.There was no patient injury, medical intervention, or extended procedure time reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Type of Device
SCALPEL, ULTRASONIC, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
Manufacturer (Section G)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
Manufacturer Contact
moira barton varty
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key5570392
MDR Text Key42984423
Report Number0001056128-2016-00038
Device Sequence Number1
Product Code NLQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133672
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/25/2016
Device Catalogue NumberHAR36RR
Device Lot Number4155314
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/25/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
-
-