Model Number 4415
Device Problems False Alarm (1013); Loose or Intermittent Connection (1371)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/28/2016
Event Type  Malfunction  
Manufacturer Narrative

(b)(4). The reported complaint was confirmed. Manufacturer technical support troubleshooting provided the following solution with the biomed. The biomed reseated the computer card and the reported issue was resolved. The unit was returned to clinical use. Per the biomed: they do their own maintenance every three months, the water was not cloudy or discolored and he had no knowledge of patient infection associated with the cooler heater. No parts will be returned to the manufacturer for evaluation. If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.

 
Event Description

It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the light emitting diode (led) lights did not come on main display when started. Start-up alarm worked as it should. As a result, an alternate device was employed. The surgical procedure was completed successfully. There was no delay, no blood loss, nor adverse consequences to the patient.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSARNS TCM II HEATING AND COOLING SYSTEM
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5590452
MDR Text Key43794578
Report Number1828100-2016-00262
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation BIOMEDICAL ENGINEER
Type of Report Initial
Report Date 04/20/2016
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received04/20/2016
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number4415
Device Catalogue Number4415
Was Device Available For Evaluation? Yes
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/28/2016
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured07/02/1997
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-