Device Problem
Device Stops Intermittently (1599)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.
If the device is returned, an evaluation shall be completed and a supplemental report will be filed.
No conclusions can be made at this time.
Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging a power (intermittent power) issue.
This complaint is being reported because the reported issue was not resolved with troubleshooting.
There was no indication that the product caused or contributed to an adverse event.
|
|
Manufacturer Narrative
|
Follow-up #1: date of submission 06/08/2016.
Device evaluation: the device has been returned and evaluated by product analysis on 05/26/2016 with the following findings: the reported date of 3/1/16 is overwritten.
Several unexpected por's observed in black box on (b)(6) 2016.
Cracked battery compartment below grip pad to case seal.
Battery cap is able to fit to maintain electrical connection.
The pump powers on correctly no power interruption was duplicated during investigation.
Tightened battery cap fully then loosened one half turn, power to pump was not interrupted.
Opened pump, no intermittent conditions found on power pcb.
Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
|