• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER Back to Search Results
Model Number MERGE HEMODYNAMICS V9.40
Device Problem Application Network Problem (2879)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/25/2016
Event Type  malfunction  
Manufacturer Narrative
Further investigation by support and the customer's it determined that the network cable connecting the hemo client pc to the network switch was not fully seated.Upon reaseating the cable, the customer was able to launch the application without issue.
 
Event Description
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the hemodynamics hemo monitor pc.On (b)(6) 2016, a customer reported to merge healthcare that they were unable to launch the hemo application on their client pc during an active case.The patient was moved to another cath lab in order to complete the procedure.The customer stated there was no adverse patient care noted in this incident.Due to the users inability to launch the hemo client during a procedure and the need for the patient to be moved, there is a potential for a delay in care that could result in harm to the patient.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge drive
hartland, WI 53029
2629123570
MDR Report Key5601143
MDR Text Key43489861
Report Number2183926-2016-00554
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMODYNAMICS V9.40
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/25/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/21/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-