• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON AQUATHERM III,ELECTRONIC HEATER; BREATHING SYSTEM HEATER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON AQUATHERM III,ELECTRONIC HEATER; BREATHING SYSTEM HEATER Back to Search Results
Catalog Number 050-14
Device Problems Failure to Power Up (1476); Smoking (1585)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Event Date 12/01/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).It is unknown if the device sample is available for evaluation.A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the product was not returned.The device history record investigation did not show issues related to complaint.A document assessment (fmea) was conducted and no changes required.Corrective actions cannot be established since it is necessary to receive the device sample to perform a proper investigation and confirm the alleged defect.At this time since the device sample is not available it is not possible to determine the source of the defect reported.Customer complaint cannot be confirmed since the device sample is not available to perform a proper investigation and determine the root cause.If the device sample becomes available at a later date, this complaint will be updated accordingly.Telelfex will continue to be monitoring trending on similar complaints.
 
Event Description
The customer alleges that the device stated to smoke.The device was turned off and would not turn on again.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and no issues were observed.Functional testing was also performed and the unit passed all tests.The unit heated up on the minimum, medium and maximum temperature setting during testing and the unit did not smoke.Based on the investigation performed, the reported complaint could not be confirmed.There were no issues found with the returned device.
 
Event Description
The customer alleges that the device stated to smoke.The device was turned off and would not turn on again.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON AQUATHERM III,ELECTRONIC HEATER
Type of Device
BREATHING SYSTEM HEATER
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key5610929
MDR Text Key43776708
Report Number3003898360-2016-00450
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeMX
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number050-14
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received06/27/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-