Catalog Number 6541-3-602
Device Problems Degraded (1153); Leak/Splash (1354); Material Integrity Problem (2978)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 04/01/2016
Event Type  malfunction  
Manufacturer Narrative
When completed, the investigation results will be submitted in a supplemental report.
 
Event Description
Triathlon knee instruments that have green rubber handles had an oily substance oozing from them.
 
Manufacturer Narrative
The device was not returned for evaluation.An event regarding santoprene handle degradation was reported.The event was not confirmed.Method & results: -device evaluation and results: could not be performed as no items associated with the event were returned or made available for identification or evaluation.-medical records received and evaluation: no patient medical records were available for review.-device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies -complaint history review: there has been one other event for the lot referenced.Conclusions: a capa was initiated because a series of complaints were received for triathlon instrumentation where green santoprene over-mold handles have been reported to be degrading, becoming sticky and unstable.This event was determined to be under the scope of the capa.Chemical and cytotoxicity tests showed the degraded santoprene is non-toxic.The ability to clean and sterilize the degraded instruments has not been compromised.
 
Event Description
Triathlon knee instruments that have green rubber handles had an oily substance oozing from them.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATELLA CALIPER
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5618712
MDR Text Key44516182
Report Number0002249697-2016-01419
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123486
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number6541-3-602
Device Lot NumberSB3H27
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/31/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/28/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-