(b)(4).The device was returned for evaluation.The homechoice device received a returned instrument testing evaluation (rite).This evaluation included functional and electrical testing of the device.Device passed rite electrical test, failed rite functional test due to heater temperature timeout.Simulated-use testing was performed.A visual inspection was performed.The reported issue of burnt odor was verified during internal inspection.The visual inspection reveal the solid state relay u2 on accomp pcb was overheated.The cause of the condition was determined to be an overheated u2 solid state relay on the accomp board.To correct the condition, the accomp board was replaced.Should additional relevant information become available, a supplemental report will be submitted.
|