• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION ALARIS® PUMP MODULE ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION ALARIS® PUMP MODULE ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 2426-0007
Device Problem Deformation Due to Compressive Stress (2889)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/07/2016
Event Type  malfunction  
Manufacturer Narrative
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
 
Event Description
The customer reported a ballooned pump segment during an infusion of normal saline at 100ml/hr.There was no patient harm.
 
Manufacturer Narrative
The customer¿s report of a ballooned segment was confirmed.Visual examination showed that the silicone segment was slightly ballooned near the upper fitment.Tactile examination confirmed that the silicone segment had been weakened in the same area.Functional testing and pressure testing replicated the ballooning.The root cause for the reported bulge in the tubing could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS® PUMP MODULE ADMINISTRATION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
stephen bilello
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key5628651
MDR Text Key44314929
Report Number9616066-2016-00582
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2426-0007
Device Catalogue Number2426-0007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/07/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-