Model Number MMT-751NAH |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Hyperglycemia (1905); Pneumonia (2011); No Code Available (3191)
|
Event Date 09/01/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
Customer reported via phone call to have been hospitalized on (b)(6) 2016 with blood glucose of 300 mg/dl.Customer was hospitalized due to high blood glucose and pneumonia.Customer was still hospitalized after five weeks.Customer was wearing insulin pump at the time of hospitalization.Customer called back on (b)(6), while still hospitalized, stating that buttons were not responding on insulin pump.Insulin pump would be replaced.
|
|
Manufacturer Narrative
|
The pump passed the functional testing.However, the pump had intermittent button response during testing due to a flattened dome switch on the up arrow button, down arrow button, esc and act buttons.The lcd connector was inspected and no anomaly was noted.No physical damage was noted during visual inspection.
|
|