MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP; OZO Back to Search Results
Model Number MMT-751NAH
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hyperglycemia (1905); Pneumonia (2011); No Code Available (3191)
Event Date 09/01/2015
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Customer reported via phone call to have been hospitalized on (b)(6) 2016 with blood glucose of 300 mg/dl.Customer was hospitalized due to high blood glucose and pneumonia.Customer was still hospitalized after five weeks.Customer was wearing insulin pump at the time of hospitalization.Customer called back on (b)(6), while still hospitalized, stating that buttons were not responding on insulin pump.Insulin pump would be replaced.
 
Manufacturer Narrative
The pump passed the functional testing.However, the pump had intermittent button response during testing due to a flattened dome switch on the up arrow button, down arrow button, esc and act buttons.The lcd connector was inspected and no anomaly was noted.No physical damage was noted during visual inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP
Type of Device
OZO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key5629375
MDR Text Key44331310
Report Number3004209178-2016-47585
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
PMA/PMN Number
P120010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 04/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-751NAH
Device Catalogue NumberMMT-751NAH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received04/28/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age57 YR
Patient Weight163
-
-