• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065751613
Device Problem Air Leak (1008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/11/2016
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that there was a bubble during a vitreoretinal procedure.The product was replaced and the procedure was completed without consequences to the patient.Additional information has been requested.A product sample was received at the manufacturing site and it is awaiting evaluation.
 
Manufacturer Narrative
The lot complaint history was reviewed; this is the first complaint for the finish goods lot and first for this issue.The device history record shows the product was released per specifications.Only the cassette with drain bag was returned for visual and functional testing.No obvious defects were observed.Lab stock components were used to replace missing items and continue testing.A console representing the current software version was used to test the sample.The ball in the check valve of the drip chamber moved freely per specification.The sample could prime and pass intraocular pressure calibration successfully.No anomalies were observed during priming.Toggling the infusion and the fluid/air exchange (f/ax) modes, fluid and air flowed from the cassette to the infusion line continuously without any bubble in various settings in all sub modes.No message code appeared on the screen.No leakage was detected from the pump elastomer or on the pump area of the fluidics module.Cleaning process was able to be performed after functional test had completed.The root cause of the customer's complaint could not be established; the returned sample met specifications.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
rita lopez
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514846
MDR Report Key5630658
MDR Text Key45097772
Report Number1644019-2016-00742
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2017
Device Catalogue Number8065751613
Device Lot Number1837881H
Other Device ID Number380657516131
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/23/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received07/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/13/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-