• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. 49" TRANSFER CARRIAGE; TRANSFER CARRIAGE TO STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. 49" TRANSFER CARRIAGE; TRANSFER CARRIAGE TO STERILIZER Back to Search Results
Device Problem Component Falling (1105)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/11/2016
Event Type  No Answer Provided  
Manufacturer Narrative
A steris service technician inspected the transfer carriage and found that the locking mechanism was slightly low.If the locking mechanism is not at the proper height for locking to the sterilizer while loading and unloading it may cause the reported event to occur.The technician adjusted the locking mechanism, tested the transfer carriage and confirmed it to be operating properly.No additional issues have been reported.
 
Event Description
The user facility reported that an employee was in the process of unloading the sterilizer and the front of the transfer carriage fell to the ground.No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
49" TRANSFER CARRIAGE
Type of Device
TRANSFER CARRIAGE TO STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX   67190
Manufacturer Contact
kathryn cadorette
5960 heisley road
mentor, OH 44060
4403927231
MDR Report Key5633879
MDR Text Key44904474
Report Number3005899764-2016-00034
Device Sequence Number1
Product Code FLE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/11/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-