Device Problems
Moisture Damage (1405); Battery Problem (2885)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas alleging a battery life issue.There was reportedly moisture in the battery compartment.There was no allegation of an adverse event associated with this complaint.This complaint is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
|
|
Manufacturer Narrative
|
Device evaluation: the pump has been returned and evaluated by product analysis on 06/03/2016 with the following findings: a visual inspection of the pump showed that the battery compartment was cracked.There was evidence of moisture contamination in the battery compartment.The returned battery cap was unable to fully tighten on to the pump.The pump was unresponsive with the returned battery cap and a test cap.The pump failed a leak test at the battery compartment.The pump cover was opened and no additional internal moisture damage was found.The pump¿s history and black box could not be downloaded due to moisture damage.
|
|