Device Problems Moisture Damage (1405); Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event Description
On (b)(6) 2016, the reporter contacted animas alleging a battery life issue.There was reportedly moisture in the battery compartment.There was no allegation of an adverse event associated with this complaint.This complaint is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
 
Manufacturer Narrative
Device evaluation: the pump has been returned and evaluated by product analysis on 06/03/2016 with the following findings: a visual inspection of the pump showed that the battery compartment was cracked.There was evidence of moisture contamination in the battery compartment.The returned battery cap was unable to fully tighten on to the pump.The pump was unresponsive with the returned battery cap and a test cap.The pump failed a leak test at the battery compartment.The pump cover was opened and no additional internal moisture damage was found.The pump¿s history and black box could not be downloaded due to moisture damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key5648000
MDR Text Key45033344
Report Number2531779-2016-09705
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 04/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Other Device ID Number011084040610006821276373816
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Device Age11 MO
Date Manufacturer Received04/27/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age9 YR
-
-