TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP, SYSTEM 1
|
Back to Search Results |
|
Model Number 801041 |
Device Problem
Positioning Problem (3009)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/20/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The field service representative (fsr) was given this information from the customer when he was at the account for preventive maintenance (pm).The fsr verified the reported complaint.He lubricated the occlusion knob/drag pellet area to free it up.Verified occlusion easier to adjust roller pump head and checked operation.The unit operated to manufacturer specifications and was returned to clinical use.No part will be returned to the manufacturer for evaluation.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the occlusion knob was hard to turn on the roller pump.The device was not changed out, as they continued to use.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|