• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C8310R
Device Problem Device Emits Odor (1425)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/19/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received and is in the process of being evaluated.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that was a burning smell coming from a homechoice device during therapy.The home patient was connected.The technical service representative initiated a swap of the device and discussed the use of continuous ambulatory peritoneal dialysis until the new machine arrived.The patient would use the pro card to program the new machine.There was no patient injury or medical intervention indicated at the time of the report.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device was received for evaluation.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Internal and external inspection was performed and no issues were noted.The homechoice device received a returned instrument testing evaluation (rite).This evaluation included functional and electrical testing of the device.The device was determined to meet functional and electrical performance specification requirements per rite testing.A simulated therapy was performed with no issues found.The reported problem of burning smell was neither verified nor duplicated.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5654159
MDR Text Key45160326
Report Number1416980-2016-08973
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/13/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C8310R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received06/03/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age45 YR
-
-