• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION ALARIS PUMP MODULE ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION ALARIS PUMP MODULE ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 2420-0007
Device Problems Stretched (1601); Deformation Due to Compressive Stress (2889)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/20/2016
Event Type  malfunction  
Manufacturer Narrative
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
 
Event Description
The customer reported that a bulge in the silicone segment of the tubing occurred while patient received long term tpn.
 
Manufacturer Narrative
The customer¿s report of bulge/balloon in the silicone segment below the upper fitment was confirmed.The set was visually inspected for kinks, incomplete bonding engagements, holes/tears in the tubing or damages to the components.During inspection it was observed that the silicone segment tubing had a small bulge and was weakened just below the upper fitment.White liquid was also observed inside the set¿s tubing.It was observed that the primary set was received without a drip chamber.Additionally observed was a knot tied in the tubing just below where the drip chamber would be connected.No other anomalies were observed on the set.Functional testing was performed; the set flowed freely and ran with no issues observed.Pressure testing confirmed a small amount of air bubbles in the silicone segment tubing when the device door was opened after infusion was completed.The root cause of the customer¿s report of ballooning silicone segment was not identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS PUMP MODULE ADMINISTRATION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
ade ajibade
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key5666673
MDR Text Key45518032
Report Number9616066-2016-00692
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2420-0007
Device Catalogue Number2420-0007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/01/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-