Device Problem
Vibration (1674)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging a audio tone/vibration (no sounds) issue.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no indication that the product caused or contributed to an adverse event.
|
|
Manufacturer Narrative
|
Follow-up #1: date of submission 04/05/2017.Device evaluation: the device has been returned and evaluated by product analysis on 03/15/2017 with the following findings: during investigation, the pump was returned with audio tone alarm inoperable.The pump was opened and a cold/cracked colder was found on the piezo.Unrelated to the original complaint, the pump casing was found to be cracked.The battery compartment was also found to be cracked.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
|